Oct 5, 2009
Endovascular Device Testing Videos
MET has access to sophisticated equipment for measuring the performance of interventional devices and vascular implants. Products tested include: heart valves, vascular implants, catheters, guide-wires, stent delivery systems, colonoscopes, endoscopes and many more.
Testing can be recorded and is both comparative and quantitative.
Typical standards tested to are: ASTM F2394-04 for preconditioning of the stent on the deployment system prior to retention testing, ISO Standard 25539-1:2003(E) Titled "Cardiovascular Implants - Endovascular Devices" and ISO/TS 15539:2000(E) Titled "Cardiovascular Implants - Endovascular Prostheses".
Standard Testing Protocols:
-
Track force - Uses the Proximal load cell to measure the force to advance a device such as a catheter, delivery system, stent system, endoprosthesis system, etc, through a tortuous anatomy with or without the aid of a guiding accessory such as a guidewire, guide catheter, guide sheath, microcatheter, etc.
-
Withdraw force - Uses the Proximal load cell to measure the force to retract a device such as a catheter, delivery system, stent system, endoprosthesis system, etc, through a tortuous anatomy with or without the aid of a guiding accessory such as a guidewire, guide catheter, guide sheath, microcatheter, etc. after accessing the intended anatomy site.
-
Push efficiency - Uses the proximal and the distal load cells to measure the amount of force the at transfered to the distal tip of a device such as a catheter, delivery system, stent system, endoprosthesis system.
- Insertion Force Measurement - Uses the proximal load cell to measures the force used to advance through the introducer sheath.
-
Guidewire Movement - Uses the proximal load cell to measure the force needed to advance a guidewire though a catheter, guide catheter or other interventional device.
-
Flexibility - Uses the proximal load cell to measure the ability of a device such as a catheter, delivery system, stent system, endoprosthesis system, etc. to advance and withdraw, with no loss of function or damage to the tortuous anatomy, over a specific bend such as 90 degrees.
-
Crossability - Uses the Proximal load cell to measure the ability of a device such as a catheter, delivery system, stent system, endoprosthesis system, etc. to cross (traverse) a given feature of a tortuous anatomy such as a lesion, stenosis, stent, etc. in the delivery path with the portion of the device required for the intended therapy.
-
Torquability - Uses the proximal and distal torque sensors to measure the amount of torque transmitted through the device such as a catheter, delivery system, stent system, endoprosthesis system, by rotating the device at a more proximal location and fixing the distal end while the device is routed through tortuous anatomy.
-
Rotational Response - Uses the proximal torque sensor and the distal rotations out fixture to measure the ability to cause axial rotation of the distal end of a device such as a catheter, delivery system, stent system, endoprosthesis system, etc by rotating the device at a more proximal location when the device is routed through a tortuous anatomy.
-
Torque to Failure - Uses the proximal torque sensor and distal torque or distal tension to measure the proximal torque and distal torque that causes failure in a device (constant length test) or to measure the proximal torque that causes failure in a device by applying constant tension (constant tension test).




