Medical And Combination Device Testing News

News

BACK
Dec 6, 2018 |
MET News, 2018,  |
A.Page

Complying with USP 661.1/661.2

When will my containers need to comply with USP <661.1> and USP <661.2>?

The new versions of these monographs will come into force on the 1st of May 2020. Old and new packaging systems will need to comply from this date, having been assessed using up to date tests. The required information includes extractable and leachable based safety assessments, biological reactivity testing and toxicological risk assessments.

One primary concern is the possibility of metal ions transferring into the formulation. As in previous versions of the methods, there is also a variety of analyses to compare the container materials to USP reference materials. These materials are cyclic olefins, polyethylene, polypropylene,

polyethylene terephthalate, and plasticized polyvinyl chloride. When other materials are employed similar analysis is required along with evidence to demonstrate toxic materials are not transferred to a patient. This means full chemical analysis of extractables accompanied by a toxicological risk analysis, although this is not specifically stated in the monograph.

Finally there is a requirement for biological reactivity/cytotoxicity testing according to USP <87> and for more invasive formulations USP <88> class testing.